Abstract
Background: Customized and standard automated insulin delivery (AID) systems for use in pregnancies of women with preexisting type 1 diabetes (T1D) are being developed and tested to achieve pregnancy appropriate continuous glucose monitoring (CGM) targets. Guidance on the use of CGM for treatment decisions during pregnancy in the United States is limited.
Methods: Ten pregnant women with preexisting T1D participated in a trial evaluating at-home use of a pregnancy-specific AID system. Seven-point self-monitoring of blood glucose (SMBG) was compared to the closest sensor glucose (Dexcom G6 CGM) value biweekly to assess safety and reliability based on the 20%/20 mg/dL criteria.
Results: All participants completed the study with 7 participants satisfying the safety and reliability criteria with a mean absolute relative difference of 10.3%. Three participants did not fulfill the criteria, mainly because the frequency of SMBG did not meet the requirements.
Conclusion: Dexcom G6 CGM is safe and accurate in the real-world setting for use in pregnant women with preexisting T1D with reduced SMBG testing as part of a pregnancy-specific AID system.
Keywords: automated insulin delivery; continuous glucose monitoring; outpatient; pregnancy; self-monitoring blood glucose; type 1 diabetes.
Published by Oxford University Press on behalf of the Endocrine Society 2024.
Similar articles
-
Performance of the Dexcom G6 Continuous Glucose Monitoring System in Pregnant Women with Diabetes.Diabetes Technol Ther. 2020 Dec;22(12):943-947. doi: 10.1089/dia.2020.0085. Diabetes Technol Ther. 2020. PMID: 32324061 Free PMC article.
-
Evaluation of the Long-Term Cost-Effectiveness of the Dexcom G6 Continuous Glucose Monitor versus Self-Monitoring of Blood Glucose in People with Type 1 Diabetes in Canada.Clinicoecon Outcomes Res. 2021 Aug 11;13:717-725. doi: 10.2147/CEOR.S304395. eCollection 2021. Clinicoecon Outcomes Res. 2021. PMID: 34408456 Free PMC article.
-
Self-measurement of Blood Glucose and Continuous Glucose Monitoring - Is There Only One Future?Eur Endocrinol. 2018 Sep;14(2):24-29. doi: 10.17925/EE.2018.14.2.24. Epub 2018 Sep 10. Eur Endocrinol. 2018. PMID: 30349591 Free PMC article. Review.
-
The effectiveness of continuous subcutaneous insulin pumps with continuous glucose monitoring in outpatient adolescents with type 1 diabetes: A systematic review.JBI Libr Syst Rev. 2012;10(42 Suppl):1-10. doi: 10.11124/jbisrir-2012-170. JBI Libr Syst Rev. 2012. PMID: 27820140
-
Continuous glucose monitoring systems for type 1 diabetes mellitus.Cochrane Database Syst Rev. 2012 Jan 18;1(1):CD008101. doi: 10.1002/14651858.CD008101.pub2. Cochrane Database Syst Rev. 2012. PMID: 22258980 Free PMC article. Review.
References
-
- Maresh MJ, Holmes VA, Patterson CC, et al. . Glycemic targets in the second and third trimester of pregnancy for women with type 1 diabetes. Diabetes Care 2015;38(1):34‐42. - PubMed
-
- Cohen O, Keidar N, Simchen M, Weisz B, Dolitsky M, Sivan E. Macrosomia in well controlled CSII treated type I diabetic pregnancy. Gynecol Endocrinol 2008;24(11):611‐613. - PubMed
-
- Ekbom P, Damm P, Feldt-Rasmussen B, Feldt-Rasmussen U, Jensen DM, Mathiesen ER. Elevated third-trimester haemoglobin A 1c predicts preterm delivery in type 1 diabetes. J Diabetes Complications 2008;22(5):297‐302. - PubMed
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous