Simultaneous R2*, R2, and R2' quantification by combining S0 estimation of the free induction decay with a single spin echo: A single acquisition method for R2 insensitive quantification of holmium-166-loaded microspheres
- PMID: 24610772
- DOI: 10.1002/mrm.25138
Simultaneous R2*, R2, and R2' quantification by combining S0 estimation of the free induction decay with a single spin echo: A single acquisition method for R2 insensitive quantification of holmium-166-loaded microspheres
Abstract
Purpose: To present a new method, S0 estimation of the free induction decay combined with a single spin echo measurement (SOFIDSE), that enables simultaneous measurements of R2*, R2, and R2' in order to quantify the local concentration of holmium microspheres (Ho-MS) for radioembolization.
Theory and methods: SOFIDSE estimates R2* and the signal magnitude at time point 0, S0, from a multigradient echo readout of the free induction decay and subsequently estimates R2 using S0 and a single spin echo, from which R2' is deducted. The method was evaluated by comparing SOFIDSE R2 values with values obtained from shifted spin echo (SSE) measurements on a phantom setup containing Ho-MS and from dual spin echo measurements on a healthy volunteer.
Results: On average, SOFIDSE showed a small overestimation of R2 values compared with SSE independent of the microsphere concentration. R2' values determined by subtraction of either SOFIDSE R2 or SSE R2 from R2* showed excellent agreement (correlation coefficient = 1; P = 9 · 10(-11)). The Ho-MS-induced R2' values obtained by SOFIDSE were insensitive to the R2 value of the tissue in which they resided.
Conclusion: SOFIDSE enables quantification of Ho-MS, in media with spatially or temporally varying R2 values, in a single acquisition.
Keywords: R2; R2*; S0 estimation; SOFIDSE; holmium microspheres; quantification.
© 2014 Wiley Periodicals, Inc.
Similar articles
-
FID sampling superior to spin-echo sampling for T2*-based quantification of holmium-loaded microspheres: theory and experiment.Magn Reson Med. 2008 Dec;60(6):1466-76. doi: 10.1002/mrm.21785. Magn Reson Med. 2008. PMID: 19026005
-
Quantification of holmium-166 loaded microspheres: estimating high local concentrations using a conventional multiple gradient echo sequence with S₀-fitting.J Magn Reson Imaging. 2012 Jun;35(6):1453-61. doi: 10.1002/jmri.23593. Epub 2012 Jan 26. J Magn Reson Imaging. 2012. PMID: 22281661
-
Reversible, irreversible and effective transverse relaxation rates in normal aging brain at 3T.Neuroimage. 2014 Jan 1;84:1032-41. doi: 10.1016/j.neuroimage.2013.08.051. Epub 2013 Sep 1. Neuroimage. 2014. PMID: 24004692
-
Value of transverse relaxometry difference methods for iron in human brain.Magn Reson Imaging. 2016 Jan;34(1):51-9. doi: 10.1016/j.mri.2015.09.002. Epub 2015 Oct 3. Magn Reson Imaging. 2016. PMID: 26435459
-
Single spin-echo proton transverse relaxometry of iron-loaded liver.NMR Biomed. 2004 Nov;17(7):446-58. doi: 10.1002/nbm.905. NMR Biomed. 2004. PMID: 15523601 Review.
Cited by
-
In vivo measurements of irreversible and reversible transverse relaxation rates in human basal ganglia at 7 T: making inferences about the microscopic and mesoscopic structure of iron and calcification deposits.NMR Biomed. 2019 Nov;32(11):e4140. doi: 10.1002/nbm.4140. Epub 2019 Jul 19. NMR Biomed. 2019. PMID: 31322331 Free PMC article.
-
Dual-pathway multi-echo sequence for simultaneous frequency and T2 mapping.J Magn Reson. 2016 Apr;265:177-87. doi: 10.1016/j.jmr.2016.01.019. Epub 2016 Feb 4. J Magn Reson. 2016. PMID: 26923150 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources