Ground test results of the micro-vibration interference for the x-ray microcalorimeter onboard XRISM
Authors:
Takashi Hasebe,
Ryuta Imamura,
Masahiro Tsujimoto,
Hisamitsu Awaki,
Meng P. Chiao,
Ryuichi Fujimoto,
Leslie S. Hartz,
Caroline A. Kilbourne,
Gary A. Sneiderman,
Yoh Takei,
Susumu Yasuda
Abstract:
Resolve is a payload hosting an X-ray microcalorimeter detector operated at 50 mK in the X-Ray Imaging and Spectroscopy Mission (XRISM). It is currently under development as part of an international collaboration and is planned to be launched in 2023. A primary technical concern is the micro-vibration interference in the sensitive microcalorimeter detector caused by the spacecraft bus components.…
▽ More
Resolve is a payload hosting an X-ray microcalorimeter detector operated at 50 mK in the X-Ray Imaging and Spectroscopy Mission (XRISM). It is currently under development as part of an international collaboration and is planned to be launched in 2023. A primary technical concern is the micro-vibration interference in the sensitive microcalorimeter detector caused by the spacecraft bus components. We conducted a series of verification tests in 2021-2022 on the ground, the results of which are reported here. We defined the micro-vibration interface between the spacecraft and the Resolve instrument. In the instrument-level test, the flight-model hardware was tested against the interface level by injecting it with micro-vibrations and evaluating the instrument response using the 50 mK stage temperature stability, ADR magnet current consumption rate, and detector noise spectra. We found strong responses when injecting micro-vibration at about 200, 380, and 610 Hz. In the former two cases, the beat between the injected frequency and cryocooler frequency harmonics were observed in the detector noise spectra. In the spacecraft-level test, the acceleration and instrument responses were measured with and without suspension of the entire spacecraft. The reaction wheels (RWs) and inertial reference units (IRUs), two major sources of micro-vibration among the bus components, were operated. In conclusion, the observed responses of Resolve are within the acceptable levels in the nominal operational range of the RWs and IRUs. There is no evidence that the resultant energy
resolution degradation is beyond the current allocation of noise budget.
△ Less
Submitted 2 March, 2023;
originally announced March 2023.
Mechanical cryocooler noise observed in the ground testing of the Resolve X-ray microcalorimeter onboard XRISM
Authors:
R. Imamura,
H. Awaki,
M. Tsujimoto,
S. Yamada,
F. S. Porter,
C. A. Kilbourne,
R. L. Kelley,
Y. Takei
Abstract:
Low-temperature detectors often use mechanical coolers as part of the cooling chain in order to reach sub-Kelvin operating temperatures. The microphonics noise caused by the mechanical coolers is a general and inherent issue for these detectors. We have observed this effect in the ground test data obtained with the Resolve instrument to be flown on the XRISM satellite. Resolve is a cryogenic X-ray…
▽ More
Low-temperature detectors often use mechanical coolers as part of the cooling chain in order to reach sub-Kelvin operating temperatures. The microphonics noise caused by the mechanical coolers is a general and inherent issue for these detectors. We have observed this effect in the ground test data obtained with the Resolve instrument to be flown on the XRISM satellite. Resolve is a cryogenic X-ray microcalorimeter spectrometer with a required energy resolution of 7 eV at 6 keV. Five mechanical coolers are used to cool from ambient temperature to about 4 K: four two-stage Stirling coolers (STC) driven nominally at 15 Hz and a Joule-Thomson cooler (JTC) driven nominally at 52 Hz. In 2019, we operated the flight-model instrument for two weeks, in which we also obtained accelerometer data inside the cryostat at a low-temperature stage (He tank). X-ray detector and accelerometer data were obtained continuously while changing the JTC drive frequency, which produced a unique data set for investigating how the vibration from the cryocoolers propagates to the detector. In the detector noise spectra, we observed harmonics of both STCs and JTC. More interestingly, we also observed the low (<20 Hz) frequency beat between the 4'th JTC and 14'th STC harmonics and the 7'th JTC and the 23--24'th STC harmonics. We present here a description and interpretation of these measurements.
△ Less
Submitted 2 March, 2023;
originally announced March 2023.