Euclid Near Infrared Spectrometer and Photometer instrument flight model presentation, performance and ground calibration results summary
Authors:
T. Maciaszek,
A. Ealet,
W. Gillard,
K. Jahnke,
R. Barbier,
E. Prieto,
W. Bon,
A. Bonnefoi,
A. Caillat,
M. Carle,
A. Costille,
F. Ducret,
C. Fabron,
B. Foulon,
J. L. Gimenez,
E. Grassi,
M. Jaquet,
D. Le Mignant,
L. Martin,
T. Pamplona,
P. Sanchez,
J. C. Clémens,
L. Caillat,
M. Niclas,
A. Secroun
, et al. (73 additional authors not shown)
Abstract:
The NISP (Near Infrared Spectrometer and Photometer) is one of the two Euclid instruments. It operates in the near-IR spectral region (950-2020nm) as a photometer and spectrometer. The instrument is composed of: a cold (135 K) optomechanical subsystem consisting of a Silicon carbide structure, an optical assembly, a filter wheel mechanism, a grism wheel mechanism, a calibration unit, and a thermal…
▽ More
The NISP (Near Infrared Spectrometer and Photometer) is one of the two Euclid instruments. It operates in the near-IR spectral region (950-2020nm) as a photometer and spectrometer. The instrument is composed of: a cold (135 K) optomechanical subsystem consisting of a Silicon carbide structure, an optical assembly, a filter wheel mechanism, a grism wheel mechanism, a calibration unit, and a thermal control system, a detection system based on a mosaic of 16 H2RG with their front-end readout electronic, and a warm electronic system (290 K) composed of a data processing / detector control unit and of an instrument control unit that interfaces with the spacecraft via a 1553 bus for command and control and via Spacewire links for science data.
This paper presents: the final architecture of the flight model instrument and subsystems, and the performance and the ground calibration measurement done at NISP level and at Euclid Payload Module level at operational cold temperature.
△ Less
Submitted 18 October, 2022;
originally announced October 2022.
MOEMS deformable mirror testing in cryo for future optical instrumentation
Authors:
Frederic Zamkotsian,
Patrick Lanzoni,
Rudy Barette,
Emmanuel Grassi,
Patrick Vors,
Michael Helmbrecht,
Franck Marchis,
Alex Teichman
Abstract:
MOEMS Deformable Mirrors (DM) are key components for next generation instruments with innovative adaptive optics systems, in existing telescopes and in the future ELTs. These DMs must perform at room temperature as well as in cryogenic and vacuum environment. Ideally, the MOEMS-DMs must be designed to operate in such environment. We present some major rules for designing / operating DMs in cryo an…
▽ More
MOEMS Deformable Mirrors (DM) are key components for next generation instruments with innovative adaptive optics systems, in existing telescopes and in the future ELTs. These DMs must perform at room temperature as well as in cryogenic and vacuum environment. Ideally, the MOEMS-DMs must be designed to operate in such environment. We present some major rules for designing / operating DMs in cryo and vacuum. We chose to use interferometry for the full characterization of these devices, including surface quality measurement in static and dynamical modes, at ambient and in vacuum/cryo. Thanks to our previous set-up developments, we placed a compact cryo-vacuum chamber designed for reaching 10-6 mbar and 160K, in front of our custom Michelson interferometer, able to measure performances of the DM at actuator/segment level as well as whole mirror level, with a lateral resolution of 2μm and a sub-nanometric z-resolution. Using this interferometric bench, we tested the Iris AO PTT111 DM: this unique and robust design uses an array of single crystalline silicon hexagonal mirrors with a pitch of 606μm, able to move in tip, tilt and piston with strokes from 5 to 7μm, and tilt angle in the range of +/-5mrad. They exhibit typically an open-loop flat surface figure as good as <20nm rms. A specific mount including electronic and opto-mechanical interfaces has been designed for fitting in the test chamber. Segment deformation, mirror shaping, open-loop operation are tested at room and cryo temperature and results are compared. The device could be operated successfully at 160K. An additional, mainly focus-like, 500 nm deformation is measured at 160K; we were able to recover the best flat in cryo by correcting the focus and local tip-tilts on some segments. Tests on DM with different mirror thicknesses (25μm and 50μm) and different coatings (silver and gold) are currently under way.
△ Less
Submitted 18 April, 2017; v1 submitted 12 April, 2017;
originally announced April 2017.