. 2019 Jul;73(7):1077-1083.
doi: 10.1038/s41430-018-0355-z. Epub 2018 Oct 22.

Lack of agreement of in vivo raw bioimpedance measurements obtained from two single and multi-frequency bioelectrical impedance devices

Affiliations

Lack of agreement of in vivo raw bioimpedance measurements obtained from two single and multi-frequency bioelectrical impedance devices

Analiza M Silva et al. Eur J Clin Nutr. 2019 Jul.

Abstract

Background: It is important for highly active individuals to accurately assess their hydration level. Bioelectrical impedance (BIA) can potentially meet these needs but its validity in active individuals is not well established.

Methods: We compared whole-body bioimpedance measurements obtained from multi-frequency bioelectrical impedance spectroscopy (BIS, Xitron 4200) at a 50 kHz frequency with those determined by a phase-sensitive single-frequency device (SF-BIA, BIA-101, RJL/Akern Systems) in two populations: active adults and elite athletes.

Results: One hundred twenty-six participants, including active males involved in recreational sports (N = 25, 20-39 yr) and elite athletes (females: N = 26, 18-35 yr; males: N = 75, 18-38 yr) participated in this study. Reactance (Xc), Resistance (R), Impedance (Z), and phase angle (PhA) were obtained by BIS and SF-BIA. Small but significant differences (R: -9.91 ± 15.09 Ω; Xc: -0.97 ± 2.56 Ω; Z: -9.96 ± 15.18 Ω; PhA: 0.12 ± 0.2°) were observed between the bioimpedance equipment in all measured variables (p < 0.05) though differences were within the devices' technical error of measurements. Device-specific values were highly (p < 0.0001) correlated [R2 ranged from 0.881 (Xc) to 0.833 (R)], but slopes and intercepts were different (p < 0.0001) from 1 and 0, respectively. Relatively large limits of agreement were observed for R (-40 to 21 Ω), Xc (-6 to 4 Ω), PhA (-0.4 to 0.5°), and impedance (-40 to 20 Ω).

Conclusion: Bioimpedance measurements from the current single- and multi-frequency devices should not be used interchangeably. The of lack of agreement between devices was observed in determining individual values of R, Xc, Z and PhA of highly active populations possibly due to methodological and biological factors.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Gatterer H, Schenk K, Laninschegg L, Schlemmer P, Lukaski H, Burtscher M. Bioimpedance identifies body fluid loss after exercise in the heat: a pilot study with body cooling. PLoS One. 2014;9:e109729. - DOI
    1. Lukaski HC. Evolution of bioimpedance: a circuitous journey from estimation of physiological function to assessment of body composition and a return to clinical research. Eur J Clin Nutr. 2013;67(Suppl 1):S2–9. - DOI
    1. Earthman CP. Body composition tools for assessment of adult malnutrition at the bedside: a tutorial on research considerations and clinical applications. JPEN J Parenter Enter Nutr. 2015;39:787–822. - DOI
    1. Nescolarde L, Yanguas J, Terricabras J, Lukaski H, Alomar X, Rosell-Ferrer J, et al. Detection of muscle gap by L-BIA in muscle injuries: clinical prognosis. Physiol Meas. 2017;38:L1–L9. - DOI
    1. Nescolarde L, Yanguas J, Lukaski H, Rodas G, Rosell-Ferrer J. Localized BIA identifies structural and pathophysiological changes in soft tissue after post-traumatic injuries in soccer players. Conf Proc: Annu Int Conf IEEE Eng Med Biol Soc IEEE Eng Med Biol Soc Annu Conf. 2014;2014:3743–6.

Publication types

LinkOut - more resources